PHF Paperless Recorder

Part Number: PHF Brand: Fuji Electric Co, Ltd
Price
Request a Quote
Stock
Call for Pricing & Availablility

This is a paperless recorder that displays measured data on the LCD in real time and stores data in CompactFlash.

The type of input such as thermocouple, resistance bulb, D.C. voltage (current), etc. can be arbitrarily set to 6 channels at the maximum.

The data stored in CompactFlash can be regenerated on the screen, and the use of supplied support software allows the data to be regenerated on a PC screen.

The data recorded in ASCII format can be directly read in a spreadsheet such as Excel, which facilitates the processing on a PC.
Features
  • Large capacity storage by CompactFlash
  • Quick search and display of past data
  • Various display capability
  • PC support software supplied as standard
  • Compact size
  • 6-point recording
  • Screen saver function
  • Ethernet function

Specifications

Accuracy

  • Clock: ± 50 ppm or Less
  • Digital Indication (K Thermocouple): ± (0.3 % + 1-Digit)
  • Digital Indication (Ni100, Cu50 Resistance Bulb): ± (0.5 % + 1-Digit)
  • Digital Indication (R Thermocouple, JPt100, Pt100, Pt50 Resistance Bulb, DC Voltage): ± (0.15 % + 1-Digit)

Ambient Temperature Range

  • With Ethernet Function: 0° to 40° C (32° to 104° F)
  • Without Ethernet Function: 0° to 50° C (32° to 122° F)

Approvals & Certifications

  • EMC EN 61326
  • IEC 61010-1

Bezel Material

  • PC-ABS

Burn-Out

  • Approximately 0.2 µA

Communication

  • Ethernet
  • FTP
  • HTTP
  • Modbus TCP/IP
  • SMTP (E-mail Client)

Consumption

  • Power: 32 VA at 100 Vac
  • Power: 42 VA at 240 Vac

Electrical Connectors

  • M3 Screw Terminal

Enclosure Finish

  • Black

Enclosure, Body Material

  • PC-ABS

Frequency

  • Effect of Power Supply Fluctuation Conditions: 47 to 63 Hz (100 Vac) [Reading Change: ± (0.2 % + 1-Digit) or Lower]
  • Power Supply, Normal Operating Condition (± 2 %): 50/60 Hz (Both Employable)

Function

  • Burn-Out
  • Input Filter
  • Scaling
  • Square Rooter
  • Subtraction

Humidity

  • Ambient: 20 to 80 % Relative Humidity
  • Transportation/Storage Conditions: 5 to 90 % Relative Humidity, Non-Condensing

Hysteresis

  • Alarm Function: 0 to 100 %

Indication

  • 5.7” STN Color LCD (320 x 240 Dots) with Backlight

Input

  • 0-5 Vdc
  • 0-500 mVdc
  • Resistance Bulb Measured Current: 1 mA
  • Thermocouple, Resistance Bulb, DC Voltage: ± 10 Vdc (Continuous) - Max.

Insulation Resistance

  • 100 megaohms or more (When Measured between Each Terminal & Ground by Using A 500 Vdc Megger)

Measuring Range

  • Resistance Bulb: -200° to 600° C (-328° to 1,112° F)
  • Thermocouple: -200° to 1,760° C (-328° to 3,200° F)

Mounting

  • Panel-Mounted (Vertical Panel)

Mounting Position

  • Backward Tilt within 30° Slant [Reading Change: ± (0.2 % + 1-Digit) or Lower]
  • Mounting Posture (Forward Tilt & Horizontal): 0°

Operating Voltages

  • 90-264 Vac

Output

  • Alarm Relay Output: 3 A at 150 Vac/240 Vac (Resistive Load), 3 A at 30 Vdc

Power Supply

  • Rated: 100-240 Vac

Pulse Range

  • DI Input Pulse Width (ON/OFF): 200 milliseconds or Longer

Resistance

  • Input (Thermocouple, DC Voltage): 1 Megaohm (At Power ON)
  • Input (Thermocouple, DC Voltage): 100 kiloohms (At Power OFF)
  • Signal Source Resistance (Resistance Bulb Input): 10 ohms/Wire or Less
  • Signal Source Resistance (Thermocouple Input): 1 Kiloohm or Less

Resistance Effect

  • Thermocouple Input: 30 µV ± 1-Digit / 100 ohms

Resolution

  • Digital Indication (0-5 Vdc & 1-5 Vdc): 1 mV
  • Digital Indication (50 mVdc Voltage): 10 µV
  • Digital Indication (500 mVdc Voltage): 100 µV
  • Digital Indication: 0.1° C (0.18° F) for Thermocouple & Resistance Bulb

RTD Type

  • Cu50
  • JPt100
  • Ni100
  • Pt100
  • Pt50

Sensor, Probe Type

  • B Thermocouple
  • E Thermocouple
  • J Thermocouple
  • K Thermocouple
  • L Thermocouple
  • N Thermocouple
  • PN Thermocouple
  • R Thermocouple
  • S Thermocouple
  • T Thermocouple
  • U Thermocouple
  • W Thermocouple

Shock

  • Transportation/Storage Conditions: 294 m/sec² or Lower (Packed State)

Storage Temperature

  • -10° to 60° C (14° to 140° F)

Switch Type

  • SPST

System Requirements

  • Windows® 2000/XP

Temperature Effect

  • Ambient (Reading Change): ± (0.3 % + 1-Digit)/10° C or Lower

Temperature Range

  • Thermocouple B: 400° to 600° C (752° to 1,112° F)
  • Thermocouples K, E, J, T, L & U: -200° to -100° C (-328° to -148° F)
  • Thermocouples R & S: 0° to 300° C (32° to 572° F)
  • Transportation: -10° to 60° C (14° to 140° F)

Time Constant

  • Input Filter Function: 0 to 900 seconds

Vibration

  • Normal Operating Condition: 10-60 Hz, 0.2 m/sec² or Less [Reading Change: ± (0.2 % + 1-Digit) or Lower]
  • Transportation/Storage Conditions: 10-60 Hz, 2.45 m/sec² or Lower

Voltage

  • Withstand Voltage [Alarm Terminal - Alarm Terminal]: 750 Vac, 1 Minute
  • Withstand Voltage [Input Terminal - Ground, Communication Terminal - Ground]: 500 Vac, 1 Minute
  • Withstand Voltage [Power Terminal - Ground, Alarm Terminal - Ground]: 2,000 Vac, 1 Minute

show more