PD9500 Precision Calibrator
Brand: Precision Digital CorporationThe PD9500 calibrator precision calibrator sources voltage, current, TC, RTD, Ohms, and pressure with a second completely isolated measurement channel for a single laboratory calibration instrument unmatched in versatility, performance, and value.
The PD9500 calibrator’s world-class performance and features are accessed through a very simple-to-use, intuitive user interface.
Precision Digital’s PD9500 calibrator is truly a “process calibration laboratory in a box.”
Features- Initial Calibration Accuracy (6 months) to 0.0025% of Rdg. (voltage)
- 1-Year Accuracy to 0.005% of Rdg. (voltage)
- Source/Read Thermocouple (13), RTD (9), Voltage, Current and Pressure (read only)
- Custom RTD and SPRT Profiles
- Nine Setpoints for Each Output Range and Type
- 24 VDC Loop Power
- Direct Keyboard Entry
- Beryllium-Copper Binding Posts Reduce Thermal EMFs
- RS-232 and IEEE-488 Remote Control
- Compatible with Fluke Met/Cal® Software
- Isolated Measurement Channel
- Selectable 250 Ohm HART™ Resistor
- Includes NIST Traceable Certificate with Calibration Data
Displaying prices & availability for Ohio, United States 43201 in USD. Change location.
Models matching criteria: 2
Call for Pricing & Availablility
Does not ship to OH, United States
Call for Pricing & Availablility
Does not ship to OH, United States
Specifications
Power Supply
- Selectable 90-140 VAC, and 210-240 VAC, <15 VA
Output Voltage
- Range: 0 to 100.000 mV, 0 to 1.00000 V, 0 to 10.0000 V, 0 to 100.000 V
- Resolution: 0 to 100 mV Range 1 µV 0 to 1 V Range 10 µV 0 to 10 V Range 100 µV 0 to 100 V Range 1 mV
Resolution (Output Voltage)
- 0 to 100 mV Range: 1 µV
- 0 to 1 V Range: 10 µV
- 0 to 10 V Range: 100 µV
- 0 to 100 V Range: 1 mV
Output Current
- Range: 0 to 100.000 mA
- Resolution: 1 µA
- Accuracy (% of rdg.): ± 0.005% ± 1 Count
- Maximum Burden: 10 V
Thermocouples
- Input/Output Types J, K, T, E, R, S, N, B, L, U, C, BP, XK
- Range: mV
- Resolution: Input 0.1 °C/°F; Output 0.01°C/°F
- Accuracy: 0.14 °C; Type J, typical
RTD
- Output Range Pt385 (100, 200, 500, 1000), Pt392, Pt3916 ((JIS), Ni120, Cu 10, YSI 400
- Input Range Pt385 (100, 200, 500, 1000), Pt392, Pt3916 (JIS), Ni120, Cu10,
- Resolution: Output: 0.01 °C/°F; Pt385-1 00, typical; Input: 0.001 °C/°F; Pt385-100, typical
- Accuracy: Output: ±0.05 °C; Pt385-100, typical; Input: 0.02 °C/°F; Pt385-100, typical
Ohms
- Output Range 5 to 4000.0 Ω
- Input (4 wire connection) Range 0 to 4000.00 Ω
Pressure
- Range: 0 to 1 inch H2 O; to 10,000 psi
Operating Temperature
- 0° to 50° C (32° to 122° F)
Storage Temperature
- -20° to 70° C (-4° to 158° F)
Display
- (2) large character 16 by 2 line alphanumeric backlit LCDs